4 pool command, 5 pool one destination command, Pool command – Eppendorf epBlue with MultiCon User Manual

Page 60: Pool one destination command

Advertising
background image

epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)

60

6.3.4

Pool

command

The

Pool

command can be used to combine liquids from several source positions in one or more destination

positions.

Abb. 6-14: Principle of the Pool command

Fig. 6-14: Principle of the

Pool

command

The windows and functions of the

Pool

command are identical with the

Sample transfer

command (see

p. 48).

If you are using the

Multiaspirate

function, liquid will be removed from the source positions until the pipette

tip is filled. The contents will then be dispensed in the destination position. In the pattern, you define which
source positions will be pooled in the destination position.

The

Number of samples

command before the

Pool

command defines the number of source positions.

Options

>

Change Tips

is set to

before asp. for next destination well

.

6.3.5

Pool one destination

command

The

Pool one destination

command can be used to combine liquids from several source positions in one

destination position. This command is a simplified

Pool

command.

The

Number of samples

command before the

Pool one destination

command defines the number of source

positions.

The

Options

>

Change Tips

option is set to

when command finished

. The pipette tips are ejected at the end of

the command.