Eppendorf epBlue with MultiCon User Manual

Page 148

Advertising
background image

epBlue - Software error messages
epBlue™ with MultiCon
English (EN)

148

0x129A

Tip too small

Reagent Transfer

: Used tip is

too small.



Use a larger tip.

0x129B

Source vessel too
small

Reagent Transfer

: Used source

vessel is too small.



Use a larger vessel.

0x12A2

Check levels in
vacuum command:
Wells in filter plate
not empty

Wells of the filter plate plug
while applying vacuum.



Follow the instructions on
the Control Panel display.

0x12A4

Labware in vacuum
manifold is not
composed of three
parts (filter plate,
frame, collection
labware)

When applying vacuum three
items must be in the vacuum
unit (filter plate above vac
frame above collection plate or
tub).



See explanation in
"Cause".

0x12A5

Well pattern of
upper and lower
rack do not fit

Vacuum unit: Well geometry
of filter plate and collection
plate must correspond to each
other (e.g., 96 well filter plate
and 96 well collection plate).
Exception: Collection plate
may be replaced by a tub.



See explanation in
"Cause".

0x12B4

Temperature unit in
location C1 not
active

In the

wait

command, the

wait

for temperature

parameter was

edited for location C1 even
though no

temperature

command for C1 had been
edited before.



Insert

temperature

command before

wait

command or delete the

wait for temperature

parameter for this location.

0x12B5

Temperature unit in
location C2 not
active

In the

wait command

, the

wait

for temperature

parameter was

edited for location C2 even
though no

temperature

command for C2 had been
edited before.



Insert

temperature

command before

wait

command or delete the

wait for temperature

parameter for this location.

0x12B6

Temperature unit in
location C3 not
active

In the

wait command

, the

wait

for temperature

parameter was

edited for location C3 even
though no

temperature

command for C3 had been
edited before.



Insert

temperature

command before

wait

command or delete the

wait for temperature

parameter for this location.

0x12B7

Temperature unit in
location C1 too hot
for labware
(temperature
command in this
method)

The temperature of location
C1 (according to a

temperature

command in this application)
is too high for the labware
placed on this location.



Select a labware item that
is more resistant to high
temperatures (e.g, made
from polypropylene or
polycarbonate).

Code

Symptom/message Cause

Remedy