Eppendorf epBlue with MultiCon User Manual

Page 149

Advertising
background image

epBlue - Software error messages

epBlue™ with MultiCon

English (EN)

149

0x12B8

Temperature unit in
location C2 too hot
for labware
(temperature
command in this
method)

The temperature of location
C2 (according to a

temperature

command in this application)
is too high for the labware
placed on this location.



Select a labware item that
is more resistant to high
temperatures (e.g., made
from polypropylene or
polycarbonate).

0x12B9

Temperature unit in
location C3 too hot
for labware
(temperature
command in this
method)

The temperature of location
C3 (according to a

temperature

command in this application)
is too high for the labware
placed on this location.



Select a labware item that
is more resistant to high
temperatures (e.g., made
from polypropylene or
polycarbonate).

0x12BA

Temperature unit in
location C1 too hot
for labware
(temperature
command in
previous method)

The temperature of location
C1 (according to a

temperature

command in a previous
application with parameter

keep temperature after method
run

set active) is too high for

the labware placed on this
location.



Select a labware item that
is more resistant to high
temperatures (e.g., made
from polypropylene or
polycarbonate).



Change previous
application to not maintain
temperature after
application run.

0x12BB

Temperature unit in
location C2 too hot
for labware
(temperature
command in
previous method)

The temperature of location
C2 (according to a

temperature

command in a previous
application with parameter

keep temperature after method
run

set active) is too high for

the labware placed on this
location.



Select a labware item that
is more resistant to high
temperatures (e.g., made
from polypropylene or
polycarbonate).



Change previous
application to not maintain
temperature after
application run.

0x12BC

Temperature unit in
location C3 too hot
for labware
(temperature
command in
previous method)

The temperature of location
C3 (according to a

temperature

command in a previous
application with parameter

keep temperature after method
run

set active) is too high for

the labware placed on this
location.



Select a labware item that
is more resistant to high
temperatures (e.g., made
from polypropylene or
polycarbonate).



Change previous
application to not maintain
temperature after
application run.

0x12D0

Parameter conflict:
Elution volume too
large for this tool

Sample transfer with

elution

from filter

option:

Volume to be aspirated is too
large for the tip used.



Select a tip large enough
for picking up the liquid as
well as the additional
volume of air to be
aspirated when using this
option.

Code

Symptom/message Cause

Remedy