epBlue - Software error messages
epBlue™ with MultiCon
English (EN)

168

0x210D

Source rack not
defined

Parameter

Source

was not

edited in the application.



See explanation in
"Cause".

0x210E

Source rack not
edited in worktable

The source rack you edited in
the procedure of the
application has been removed
from the worktable.



Edit the rack that you
programmed in the
procedure as

Source

in the

worktable, or edit a
different source rack in the
application.

0x210F

Source rack not
selected in the
labware file window

The source rack you edited in
the application is not selected
or removed in the labware file
window and therefore is not
available for programming.



Select the rack in the
labware file window or edit
a different rack in the
application.

0x2110

Destination rack
not defined

Parameter

Source

was not

edited in the application.



See explanation in
"Cause".

0x2111

Destination rack
not edited in
worktable

Destination rack was edited in
the procedure of the
application, but it was not
edited in the worktable.



Edit the rack that you
programmed in the
procedure as

Destination

in

the worktable or edit a
different destination rack
in the application.

0x2112

Destination rack
not selected in the
Labware File
Window

The destination rack you
edited in the application is not
selected or removed in the
labware file window and
therefore is not available for
programming.



Select the rack in the
labware file window or edit
a different rack in the
application.

0x2113

Pattern not defined Parameter

Pattern

was not

edited in the application.



See explanation in
"Cause".

0x2114

Loading error
(invalid entry in
pattern)

Normally a system error; but
may also be caused by editing
a pattern without destination
positions; or:
File damaged.



Edit a pattern with source
and destination positions.

In other cases:



Call local Eppendorf
Service.

0x211B

Liquid type not
defined

Parameter

Liquid Type

was not

edited in the application.



Choose a "Liquid Type" for
the Liquid Handling
Command.

Code

Symptom/message Cause

Remedy