epBlue Application Editor - Creating and editing applications

epBlue™ with MultiCon

English (EN)

51

6.3.1.3

Sample transfer

>

Mix

command

The

Mix

command is also available for mixing (see p. 63).

Abb. 6-10: Sample transfer > Mix

Fig. 6-10:

Sample transfer

>

Mix

Reuse tips

Returned pipette tips will be reused.
The pipette tips will be reused within an application for the same source vessel.
Is only executed if the

restore tips

option was selected in a previous command

and the restored tips are available.
The

Reuse tips

checkbox is only available with specific settings of the

Change

Tip

option.

Information on the aspiration and transfer of liquids can be found in the hardware operating
manual.

Mix before aspirating

Mix before each aspiration. Set the parameters following activation.
If you transfer a larger volume than the maximum volume of the pipette, the
liquid will be mixed before the first aspiration.

Mix after dispensing

Mix after each dispensing. Set the parameters following activation. Available
for

Transfer type

>

Pipette

.

If you transfer a larger volume than the maximum volume of the pipette, the
liquid will be mixed after the last aspiration.

No. of Cycles

Number of mixing cycles.

Speed

Mixing frequency. If no entries were made for

Speed

, the defined aspiration

speed of the liquid type will be used.

Volume

Mixing volume

Fixed height

Position of the pipette tip on the Z-axis. Fixed aspiration level and dispensing
level. Only for levels below the working volume.

Aspiration

Distance between the pipette tip and base of the vessel during aspiration. If
0

mm is entered, the distance will be corrected to 2

mm. The correction is

dependent on the bottom tolerance and type of labware.