11 temperature command, Temperature command – Eppendorf epBlue with MultiCon User Manual

Page 65

Advertising
background image

epBlue Application Editor - Creating and editing applications

epBlue™ with MultiCon

English (EN)

65

• Before the

Dilute

command: number of samples to be diluted. The number of dilution steps is defined by

the pattern.

• Before the

Mix

command: number of source positions in which liquid is mixed.

• With an eight-channel dispensing tool: for values between 1

8, 8 samples are processed. For values

9

- 16, 16 samples are processed, etc.

6.3.11

Temperature

command

With the

Temperature

command you can temper the labware in a thermal module. The temperature control

stops when the epBlue is exited.

Abb. 6-19: Temperature window

Fig. 6-19:

Temperature

window

The command

Temperature

only is available for devices with thermal module.

Location

drop-down list

Select a thermal module.

Activate the

Temperature on

Activate the thermal module. Activated thermal
modules will be switched off when you uncheck
the checkbox.

Activate the

Keep Temperature after application

run

Activate the thermal module for the period after
the application has exited. This function can be
used to cool samples after the application run has
been completed.

Temperature

field

Set the set temperature of the thermal module.
0

°C – 110

°C

range.