Eppendorf epBlue with MultiCon User Manual

Page 125

Advertising
background image

epBlue settings - Checking components and making the basic settings

epBlue™ with MultiCon

English (EN)

125

1. Define the speed and temperature in the

Parameter

area.

2. To check the mixing function, press the

Start

button.

The thermomixer mixes until the time is elapsed or until the test has been exited using the

Stop

button.

11.7.2

Checking the temperature control of the thermomixer

Abb. 11-7: Thermomixer temperature Control function

Fig. 11-7:

Thermomixer temperature Control

function

RPM

field

Speed of the thermomixer in rpm.

Time

field

Duration of the mixing process.

Remaining time

field

Remaining time in the mixing process.

Start

button

Stop

button

Desired Temperature

field

Set temperature of the thermomixer.

Actual Temperature

field

Actual temperature of the thermomixer.

Actual current

field

Current strength via the thermal module.

Time

field

Mixing time in min.

Remain time

field

Time until the end of the mixing process
(in min).

Temp reached after

field

Time until the set temperature is reached.

New Temperature

field

Enter a new set temperature. 4

°C–

95

°C

range.

New time

field

Enter a new mixing time. 0 s

- 99 min range.

Start

button

Start the test.

Fan on

button

Switch on the fan.

Advertising