Eppendorf epBlue with MultiCon User Manual

Page 161

Advertising
background image

epBlue - Software error messages

epBlue™ with MultiCon

English (EN)

161

0x1A22, 0x1A32,
0x1A42, 0x1A52

No source or
destination selected

Edit mode/parameter in the
affected command:
Source or destination rack has
not been selected.



Select source or
destination, respectively.

0x1A40, 0x1A41,
0x1A50

8-channel tool
cannot be used for
this source rack

Edit mode/parameter in the
affected command:
Source rack does not fit the
8-channel tool (e.g., 24-well
plate or tube rack with
24

positions).



Choose another rack or
another tool.

0x1A61

8-channel tool
cannot be used for
this rack

Edit mode/parameter in
command

Mix

:

Rack does not fit the
8-channel tool (e.g., 24-well
plate or tube rack with
24

positions).



Choose another rack or
another tool.

0x1A62

No rack selected

Edit mode/parameter in
command

Mix

:

Rack has not been selected.



Select rack, respectively.

0x1A65

Invalid entry for
speed (0.2 ... 110)

Edit mode/parameter in
command

Mix

:

A value beyond the allowed
range has been entered for the
parameter

Speed

.



Enter a value between
0.2 and 110

mm/sec.

0x1A70

This position is
already occupied

Edit mode/pattern:
When editing the pattern you
have tried to select a certain
position that is already
occupied.



Follow the direction of the
edited pattern and move to
a different position.

0x1A76

8-channel tool
cannot be used for
this module rack

Edit mode/pattern:
Rack does not fit the
8-channel tool (e.g.,

Tubs +

Modules (equip) + Holders

combination with positions all
less than 8 in modules).



Choose another rack or
another tool.

0x1A77

No module rack or
tubes found

Edit mode/pattern:
Rack does not have any
positions (e.g.,

Tubs + Modules

(equip) + Holders

-combination

with positions all less than 1 in
Modules).



Choose another rack.

Code

Symptom/message Cause

Remedy

Advertising