Eppendorf epBlue with MultiCon User Manual

Page 151

Advertising
background image

epBlue - Software error messages

epBlue™ with MultiCon

English (EN)

151

0x12F7

Waiting for
thermomixer

The procedure is waiting for
the thermomixer.

Wait until thermomixer
function has ended.

0x12F8

The selected
mixing speed is not
possible with this
labware

The mixing speed is not
allowed for the selected
labware.

Select another labware or
mixing speed.

0x1500

Too big vessel
index in location: ...

A tube is to be accessed for
which the index is greater
than the number of tubes on
the plate/rack/holder.



Error during creation of
the application.

0x1503

Tool not placed ...

Optical sensor:
A tool that is needed for the
application run was not found
on the worktable.



Put the tool on the tool
holder of the worktable.

0x1504

<rack name> is not
accessible for tools
in location ...

Rack is a lower part of a
labware stack. Therefore, the
tool has no access.



Change application so that
the rack is accessible.

0x1509

Liquid volume too
large for vessel in
location:...

Total volume supplied in a
source vessel is larger than
needed or larger than vessel.



Supply less volume in the
vessel.



Change application.



When verifying the total
volume needed for the
source or destination, take
into account the additional
aspirated volume when
using multidispense mode.

0x150A

Liquid volume too
low for vessel in
location: ...

Total volume supplied by the
user in a source vessel is
smaller than the volume
needed for a sample transfer,
reagent transfer or mix
command (total volume =
volume to be aspirated +
remaining volume for this
vessel + (in case of
multidispense mode:)
additional aspirated volume.



Calculate the total volume
for the source or
destination vessel and
select a suitable vessel. For
additional aspirated
volume with multidispense
mode, refer to manual.



Consider that the software
may calculate higher
remaining volumes in
some cases to avoid
crashes.



Set

Detect volumes

to

off

for racks that are in park
positions at the beginning
of the procedure.

0x150B

Optical sensor:
Liquid volume too
low in location: ...

See above (error 0x150A).



See above (error 0x150A).

Code

Symptom/message Cause

Remedy