epBlue - Software error messages

epBlue™ with MultiCon

English (EN)

155

0x152A

Gripper could not
be found

See error message.



Provide the gripper on the
worktable at its proper
position.

0x152D

Tip too short
Select other tips or
vessels in the
method.

Tip does not reach the liquid
level at the beginning or
during the course of the liquid
handling command.



Select other tips or vessels
in the application.

0x152F

Labware stack too
high in location:
xxx
Maximum pieces
which may be piled:
5

A maximum of 5 racks can be
piled in a location. After a

transport

command, the

labware stack would be too
high.



Change application.

0x1530

Dispensing in pile
not allowed in
location: xxx

See error message.



Change application.

0x1531

Transport of this
pile not allowed in
location: xxx

Do not transport a stack.



Change application.

0x1581

Optical sensor:
Liquid level could
not be detected in
location: ...

Error in level detection.



Repeat measurement.

0x1602

Program too long

Maximum size of steps is
reached.



Delete commands that are
not necessary before insert
new commands.

0x1700

Liquid volume too
low for vessel in
location: ...

Total volume supplied by the
user in a source vessel is
smaller than the volume
needed for a sample transfer,
reagent transfer or mix
command (total volume =
volume to be aspirated +
remaining volume for this
vessel + (in case of
multidispense mode:)
additional aspirated volume.



Calculate the total volume
for the source or
destination vessel and
select a suitable vessel. For
additional aspirated
volume with multidispense
mode, refer to manual.



Consider that the software
may calculate higher
remaining volumes in
some cases to avoid
crashes.



Set

Detect volumes

to

off

for racks that are in park
positions at the beginning
of the procedure.

Code

Symptom/message Cause

Remedy