6.3.15

User

intervention

command

The

User

intervention

command is used to interrupt an application. The application will be continued when

the user confirms a message.

You can use the

User

intervention

command to empty the waste container manually.

Abb. 6-24: User Invention window

Fig. 6-24:

User Invention

window

6.3.16

Wait

command

The

Wait

command is used to create a waiting period. The application will be continued after the waiting

period. The command can be used to take into account incubation periods between the addition of
reagents.

Do not change the filling volume during the interruption. If you need to change the filling
volume, use several applications to do so.
Do not change the equipping of the worktables.

Comment

window

The comment appears when the command is
executed.

Alarm

checkbox

An acoustic signal sounds if the command is
executed.