epBlue - Software error messages

epBlue™ with MultiCon

English (EN)

159

0x1A10 8-channel

tool

cannot be used for
this source rack

Edit mode/parameter in
command

Sample Transfer

:

Source rack does not fit to
8-channel-tool (e.g.:
24-well-plate or tube rack with
24 positions).



Choose another rack or
another tool.

0x1A11

8-channel tool
cannot be used for
this destination
rack.

Edit mode/parameter in
command

Sample Transfer

:

Destination rack does not fit
the 8-channel tool (e.g.,
24-well plate or tube rack with
24

positions).



Choose another rack or
another tool.

0x1A12

No source or
destination selected

Edit mode/parameter in
command

Sample Transfer

:

Source or destination rack has
not been selected.



Select source or
destination, respectively.

0x1A15

Invalid entry for
movement blow
(0 ... 100)

Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter

Movement Blow

.



Enter a value between
0 and 100%.

0x1A15

Invalid entry for
movement blow
(0 ... 100)

Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter

Movement Blow

.



Enter a value between
0 and 100%.

0x1A16

Invalid entry for
delay blow
(0 ... 9999)

Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter

Delay Blow

.



Enter a value between
0 and 9999

msec.

0x1A16

Invalid entry for
delay blow
(0 ... 9999)

Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the

Delay Blow

parameter.



Enter a value between
0 and 9999

msec.

0x1A17

Invalid entry for
speed aspiration
(0.2 ... 110)

Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter

Speed Aspiration

.



Enter a value between
0.2 and 110

mm/sec.

Code

Symptom/message Cause

Remedy