epBlue - Software error messages
epBlue™ with MultiCon
English (EN)

146

0x0C02

Volume too small
for this tool

Volume to be dispensed is too
small for the selected tool.
Possible causes:

• Errors in tool files.
• Errors in liquid type files.



Call local Eppendorf
Application.

0x0C08

Tool dimension
unknown

Tool dimension values
unknown. Labware outdated
or corrupt.



Make sure all labware is of
the latest version.

0x120A

Program aborted by
user

User pressed the

Abort

button

during program run.



Error message was
information for the user
that he aborted the
program.

0x1221

The hood was
opened while the
program was
stopped

See error message.



Close hood.

0x1222

Transfer allowance
was prematurely
deactivated during
program
initialization

See error message.



Start program again.

0x1281

Vacuum not
reached

The vacuum chamber or rack
is leaky or no rack is inserted.



Insert another rack.



In other cases, call local
Eppendorf Service.

0x1283

Vacuum not ready

No vacuum chamber available
or vacuum chamber not ready.



Integrate or connect
vacuum chamber, or:



Call local Eppendorf
Service.

0x128B

Carrier: steps lost
in x

• Carrier collision.
• Carrier mechanics out of

order.



Ensure that the worktable
equipment in epBlue is
identical to the equipment
in epMotion, then restart
the application.



If the error persists, call
Eppendorf Service.

0x128C

Carrier: final
position in y not
found

• Problems in carrier

movement in y-axis
(sluggish movement or no
movement at all).

• Light barrier for carrier in

y-axis defective.



Call Eppendorf Service.

Code

Symptom/message Cause

Remedy