Option is set (see, P. 54) – Eppendorf epBlue with MultiCon User Manual

Page 54

Advertising
background image

epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)

54

6.3.1.5 Changing the liquid type

Abb. 6-11: Sample transfer > Liquid types

Fig. 6-11:

Sample transfer

>

Liquid types



To change the liquid type, check the

Change parameter

checkbox.



Enter the aspiration speed of the liquid in the

Aspiration speed

field: 0.2

mm/s – 110

mm/s.

For viscous solutions and large aspiration volumes, select a low aspiration speed. The liquid aspiration
will be delayed. The liquid aspiration must be ended before the pipette tip withdraws from the liquid.
Select the low aspiration speed during phase separations or to prevent the distribution of sediments or
particles.

Speed_xl

Demineralized
water.

1

µL

3

µL

• Mixing at high dispensing

speed.

• High dispensing speed poses a

high risk of contamination for
small vessels.

• Good mixing in DWP 96 with

750

µL aspiration and 750

µL

dispensing.

Speed_xs

Demineralized
water.

1

µL

3

µL

• For slow aspiration from filter

plates.

• Very low aspiration speed.

Sediment will not be
dispersed.

Water

Demineralized
water.

1

µL

3

µL

• Recommended for most

applications.

• Technical data on the

systematic and random errors
were determined using this
liquid type.

Changes to the liquid type have effects on the accuracy of pipetting. Changing the presetting
can increase the dispensing deviations.

Liquid type

Optimized for

50 µL tips:
pipetting as of

50 µL tips:
dispensing as of

Remarks