0x2026

Bottom tolerance
too small

Bottom tolerance too small.



Use a bigger value.

0x2101

Tool not defined

Parameter

Pipet. Tool

was not

edited in the application.



See explanation in
"Cause".

0x2104

Tips not edited in
worktable/
procedure

Tips were edited in the
procedure of the application,
but they were not edited in the
worktable (e.g., filter tips <->
tips without filter).



Edit the tips that you
programmed in the
procedure in the
worktable.

0x2105

Parameter conflict:
Start volume
greater than filling
volume of source
tube or well

The

Volume

and

Source

parameters of the source
vessel do not match (

Volume

is

higher than the maximum
filling volume of the source
vessel).



Edit

Volume

and

Source

in

the worktable so that

Volume

is covered by the

maximum filling volume of
the source vessel.

0x2106

Parameter conflict:
Start volume
greater than filling
volume of
destination tube or
well

The

Volume

and

Destination

parameters of the destination
vessel do not match (

Volume

is

higher than the maximum
filling volume of the
destination vessel).



Edit

Volume

and

Destination

in the

worktable so that

Volume

is

covered by the maximum
filling volume of the
destination vessel.

0x2107

Volume not defined Parameter

Volume

was not

edited in the application.



See explanation in
"Cause".

0x2108

Parameter conflict:
Volume too small
for this tool

The

Volume

and

Pipet. Tool

parameters of the application
do not match (

Volume

is

smaller than the lower limit of
the tool volume range).



Edit

Volume

and

Pipet. Tool

so that

Volume

is covered

by the volume range of the
pipette tool.

0x2109

Parameter conflict:
Volume too large
for this tool

The

Volume

and

Pipet. Tool

parameters of the application
do not match (

Volume

is

higher than the upper limit of
the tool volume range).



Edit

Volume

and

Pipet. Tool

so that

Volume

is covered

by the volume range of the
pipette tool.

0x210A

Parameter conflict:
volume greater
than filling volume
of source tube or
well

The

Volume

and

Source

parameters of the application
do not match (

Volume

is

higher than the maximum
filling volume of the source
vessel).



Edit

Volume

and

Source

so

that

Volume

is covered by

the maximum filling
volume of the source
vessel.

0x210B

Parameter conflict:
volume greater
than filling volume
of destination tube
or well

The

Volume

and

Destination

parameters of the application
do not match (

Volume

is

higher than the maximum
filling volume of the
destination vessel).



Edit

Volume

and

Destination

so that

Volume

is covered by the
maximum filling volume of
the destination vessel.

Code

Symptom/message Cause

Remedy