epBlue - Software error messages
epBlue™ with MultiCon
English (EN)

156

0x1701

Liquid volume too
large for vessel in
location:...

Total volume supplied in a
source vessel is larger than
needed or larger than vessel.



Provide less volume in the
vessel.



Change application.



When verifying the total
volume needed for the
source or destination, take
into account the additional
aspirated volume when
using multidispense mode.

epMotion 5070 only:



Set

Detect volumes

to

off

for racks that are in park
positions at the beginning
of the procedure.

0x1904

The following
labware has been
deleted: ...

Edit mode:
The worktable was changed
after an application had been
programmed; thus, the
labware defined in a command
is no longer available.



Change the

source

or

destination

in the

parameter of the respective
command in order to
match the worktable. In
this case the pattern also
has to be re-edited.



The labware has to be
reprogrammed in the
worktable.

0x1908

The method was
written with a
newer program
structure. If you
want to edit this
method, you need
to update your
software.

See Error message.



Update your software, or:



Call local Eppendorf
Service.

0x190D

The following
labware is not
selected in the
Labware File
window: ...

Edit mode/worktable:
The chosen labware is not
available in the labware
collection that was selected
for your lab. Possible cause for
this error message: The
labware has been deselected
in the labware file window.



Select the respective
labware in the Labware
File window. You need to
have the appropriate user
rights. If you do not have
the necessary user rights
ask your administrator.

Code

Symptom/message Cause

Remedy