epBlue - Software error messages

epBlue™ with MultiCon

English (EN)

171

0x2127

Mixing volume in
destination not
defined

Parameter

Volume

in a mix

procedure for destination
vessels was not edited in the
application (mix procedure as
part of a liquid transfer
command via parameter

Options

).



See explanation in
"Cause".

0x2128

Parameter conflict:
mixing volume in
destination too
large for this tool

The

Volume

and

Pipet. Tool

parameters in a mix procedure
for destination vessels do not
match (

Volume

is higher than

the upper limit of the tool
volume range) (mix procedure
as part of a liquid transfer
command via parameter

Options).



Edit

Volume

and

Pipet. Tool

so that

Volume

is within

the pipette tool's volume
range.

0x2129

Parameter conflict:
mixing volume in
destination too
small for this tool

The

Volume

and

Pipet. Tool

in

a mix procedure for
destination vessels do not
match (

Volume

is lower than

the minimum allowed volume)
(mix procedure as part of a
liquid transfer command via
parameter

Options

).



Edit

Volume

and

Pipet. Tool

so that

Volume

is within

the volume range of the
pipet tool.

0x212A

Parameter conflict:
mixing volume in
destination greater
than filling volume
of destination tube
or well

The

Volume

and

Destination

parameters in a mix procedure
for destination vessels do not
match (

Volume

is higher than

the maximum filling volume of
the destination vessel) (mix
procedure as part of a liquid
transfer command via
parameter

Options

).



Edit

Volume

and

Destination

so that

Volume

is within the maximum
filling volume of the
destination vessel.

0x212C

Parameter conflict:
mix after dispense
not allowed in
multidispense
mode

When the

transfer type

parameter is set to

multidispense

, the

mix after

dispense

parameter cannot be

edited for this command.



Change parameter

transfer

type

to

pipette

, or:



Omit the mixing step; in
this case you could also
edit another mixing step as
a new command (

Mix

),

which would be performed
after the previous
command of the procedure
has ended.

Code

Symptom/message Cause

Remedy

Advertising