Device calibration logs and records, Troubleshooting – Det-Tronics EQP Fire and Gas Detection/Releasing System User Manual

Page 108

Advertising
background image

14.1

95-8533

5-6

CALIBRATION ALGORITHM G
FOR DCUs wITH POINTwATCH OR DUCTwATCH

Routine Calibration
1. Apply the zero gas.

2. Activate the reed switch for at least 4 seconds.

(The red LED flashes for 3 seconds while the

switch is activated.)

3. The calibrate LED flashes, indicating it is ready

for the zero input.

4. When a steady zero reading is obtained, the

communication module records the uncalibrated

value in the calibration log and calibrates the zero

value during this time. The LED stays on steady.

5. Apply calibration gas. (The calibrate LED flashes

when the sensor detects gas.)

6. When the sensor input has been stable for 30

seconds, the communication module records the

uncalibrated value in the calibration log and

calibrates the span value.

7. The calibrate LED stays on steady.

8. Remove the calibration gas.

9. The communication module waits until the sensor

input drops below 4% full scale.

10. The calibration is complete. (The calibrate LED

turns off.)

Note

Calibration is aborted if not complete within 12
minutes. If not completed, the detector will revert
back to the previous calibration values. the red
LeD will flash and the calibration will be logged
as aborted.

Sensor Replacement

wARNING!

the hazardous area must be de-classified prior
to removing a junction box cover with power
applied.

1. Remove power from the DCU and PointWatch/

DuctWatch unit. Replace the PointWatch unit.

Apply power. Press the Sensor Replacement

Switch on the communication module for

approxImately 1 second.

Note

Allow at least 10 minutes for the sensor to warm
up.

Note

Pressing the Sensor Replacement Switch
prevents the communication module from
generating a fault signal when the input drops to
zero.

Note

the calibration will not be aborted if the
calibration procedure is not completed within 12
minutes.

2. Apply zero gas.

3. The calibrate LED flashes, indicating that it is

ready for the zero input.

4. Continue from step 4 of the PointWatch/

DuctWatch routine calibration procedure

described above.

DEVICE CALIBRATION LOGS AND

RECORDS

The DCU keeps a calibration log in non-volatile

memory that can be used by the operator to evaluate

the remaining life of some sensors. This log includes

the zero, span, date and time for each successful

calibration. An aborted calibration is indicated by

zeros in the zero and span values. The calibration log

is cleared when the sensor replacement switch is

pressed and the calibration is successfully

completed.

The initial calibration is logged in position one, where

it remains for the life of the sensor. If more than 8

calibrations are performed without the sensor

replacement switch being pressed, the newest

calibration data will replace the second oldest so that

the initial calibration data can be saved. The old

calibration data will be lost. This feature enables

sensor sensitivity trending to aid in maintenance or

troubleshooting.

The analog value for the sensor is represented in raw

analog-to-digital counts 0 to 4095, where 0 represents

0 ma and 4095 represents 24 ma.

TROUBLESHOOTING

Tables 5-1 and 5-2 are provided to assist in locating

the source of a system problem.

Advertising