2 setting the filter (pen model), 2 setting the filter (pen model) -2, 2 settng the flter (pen model) – Yokogawa µR10000 User Manual

Page 116

Advertising
background image

IM 04P01B01-01E

6-2

6.2 Settng the Flter (Pen Model)

This section explains the details of setting a filter on the measurement channels on the
pen model.

Procedure

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

2.

Press the

key or

SHIFT +

key to select

Aux and then press the key.

3.

Press the

key or

SHIFT +

key to select

Flter and then press the key.

4.

Set each item and press the key.

Use the

key or

SHIFT +

key to select values.

For the procedure on how to enter values or characters, see section 4.2.
If you press the

ESC key in the middle of the operation, the settings entered up to

then are cancelled, and the display returns to a higher level menu.

5.

When the

Settng complete screen appears, do either of the following:

Press the key to set other channels.
If you are done, press the

ESC key.

6.

Hold down the

MENU

key for 3 seconds to return to Operation mode.

Set the channel range.

Set the filter time constant.

The new setting takes effect.

ESC/?

First channel Last channel

Set=Aux

Aux=Filter

CH=01-01

Resp.Time=Off

01-01 Channel

Setting complete

Explanaton

Flter Tme Constant

The time constant is selectable from 2 s, 5 s, and 10 s. Select Off when not using the
filter.

<Related Topcs> Enabling the filter function: Section 7.12

Advertising