2.5 Power Supply Wrng

Precautons to Be Taken Whle Wrng the Power Supply

Make sure to follow the warnings below when wiring the power supply. To prevent electric

shock and damage to the recorder, observe the following warnings.

WARNING

Panel Mount Type

• To prevent electric shock when wiring, ensure the main power supply is turned

OFF.

• To prevent the possibility of fire, use 600 V PVC insulated wire (AWG 20 to 16)

or an equivalent wire for power wiring.

• Make sure to earth ground the protective earth terminal through a grounding

resistance less than 100 Ω before turning ON the power.

• Use crimp-on lugs (designed for 4 mm screws) for power and ground wiring

termination.

• To prevent electric shock, make sure to close the transparent cover for the

power supply wires.

• Make sure to provide a power switch (double-pole type) on the power supply line

in order to separate the recorder from the main power supply. Put an indication
on this switch as the breaker on the power supply line for the recorder and
indications of ON and OFF.

Switch specifications

Rated power current: 1 A or more (other than /P1), 3 A or more (/P1 option)
Rated rush current: 60 A or more (other than /P1), 70 A or more (/P1 option)
Complies with IEC 60947-1, 3.

• Connect a fuse in the power supply line. Use a fuse approved by CSA (for the

use in North America) or VDE (for the use in Europe).
2 A to 15 A (other than /P1), 4 A to 15 A (/P1 option)

• Do not add a switch or fuse to the ground line.
Portable Type (/H5x Opton)

• Ensure that the source voltage matches the rated power supply voltage of the

instrument before connecting the power cord.

• Connect the power cord after checking that the power switch of the portable type

is turned OFF.

• To prevent electric shock or fire, be sure to use the power cord for the portable

type supplied by YOKOGAWA.

• Make sure to connect protective earth grounding to prevent electric shock.

Connect the power cord of the portable type to a three-prong power outlet
equipped with a protective earth terminal.

• Do not use an extension cord that does not have a protective grounding wire.

The protective features of the instrument will be rendered ineffective.

Use a power supply that meets the following conditions:

Item

Power Supply Specfcatons

(Optons Other Than /P1)

(/P1 Opton)

Rated supply voltage

100 to 240 VAC

24V DC/AC

Allowable power supply voltage range

90 to 132/180 to 264 VAC

21.6 V to 26.4 VDC/AC

Rated power supply frequency

50/60 Hz

50/60 Hz (for AC)

Allowable power supply frequency range 50/60 Hz ± 2%

50/60 Hz ± 2% (for AC)

Maximum power consumption

40 VA

25 VA (for DC) or 35 VA (for AC)

Note

Do not use a supply voltage in the range 132 to 180 VAC, as this may have adverse effects on
the measurement accuracy.