Operaton errors, Communcaton errors – Yokogawa µR10000 User Manual

Page 214

Advertising
background image

IM 04P01B01-01E

10-2

10.1 A Lst of Error Messages

Code Message

Explanaton/Countermeasures

64

MATH upper and lower span values are equal.

This is not allowed.

70

MATH constant description is incorrect.

Incorrect computation constant syntax.

71

The range of the MATH constant is exceeded.

72

MATH channel is turned off

81

All space or ‘quit’ string cannot be specified.

86

The key-lock release password is incorrect.

Enter the correct password.

87

This key is locked.

91

Password is incorrect.

Enter the correct password.

100

IP address doesn’t belong to class A, B, or C.

101

The result of the masked IP address is all 0s or 1s.

102

SUBNET mask is incorrect.

Set a correct subnet mask.

103

The net part of default gateway is not equal to

that of IP address.

Set the correct default gateway.

151

This action is invalid during calculation.

160

This action is invalid during chart end.

161

This action is invalid during pen hold.

162

Cannot set an number for a skipped data.

163

This action is invalid during record.

164

This action is invalid during manual printing.

165

This action is invalid during list printing.

166

This action is invalid during setup list printing.

167

This action is invalid during chart feed.

169

This action is invalid during ribbon hold.

170

This action is invalid during priority remote record.

Recording stop via communications or key operation cannot be

accepted under the current settings. Perform the Stop using the

remote control function.

171

This action is invalid during batch.

Cannot enter Basic Setting mode during header printout (until

End printout is finished).

Operaton Errors

Code Message

Explanaton/Countermeasures

232

There is no available data.

There is no data for periodic printout or data for calculating

TLOG when the timer expired.

Communcaton Errors

Code Message

Explanaton/Countermeasures

300

Command is too long.

301

Too many number of commands delimited with ’;’.

Keep the number of commands separated by sub delimiters

under 10.

302

This command has not been defined.

303

Data request command can not be enumerated with

sub-delimiter.

350

Command is not permitted to the current user level.

351

This command cannot be specified in the current

mode.

352

The option is not installed.

353

This command cannot be specified in the current

setting.

354

This command is not available during calculation.

390

Command error.

391

Delimiter error.

392

Parameter error.

393

No permission.

394

No such connection.

395

Use “quit” to close this connection.

Attempted to disconnect its own connection.

396

Failed to disconnect.

397

No TCP control block.

The control block of the specified connection cannot be found.

400

Input username.

401

Input password.

Advertising