IM 04P01B01-01E

6-14

Explanaton

Settng Correcton Ponts (Measured Values) and Correcton Values

• Using revise values

Calibration point

B4

A3

A2

A1

A4

B3

B2

B1

Scale value

Correction

value

Measured

value

Correction using revise values

Correction point (measured value) = A

Correction value = B – A

• Using absolute values

Calibration point

B4

A3

A2

A1

A4

B3

B2

B1

Scale value

Correction

value

Measured

value

Correction using absolute values

Correction point (measured value) = A

Correction value = B

Example of a two-pont correcton

Calbraton

Pont

Corrected

Value

Measured

value (A)

Correcton Value

Revse Value

(B – A)

Absolute

Value (B)

1

10.0°C

9.8°C

0.2°C

10.0°C

2

90.0°C

90.5°C

– 0.5°C

90.0°C

Setup Condtons of Measured and Correcton Values

• Setting range of measured and correction values when using revise values
A1 < A2 ≤ A3 ≤ ... ≤ A16
(A1 + B1) < (A2 + B2) < (A3 + B3) < ... < (A16 + B16)
A and B are within the measurable range or scaling range of the range type*

* Within the scaling range: –5% to 105% of the scale value or –20000 to 30000 (the decimal

place is the same as the setting for the scale value)

• Setting range of measured and correction values when using absolute values
A1 < A2 ≤ A3 ≤ ... ≤ A16
B1 < B2 < B3 < ... < B16
A and B are within the measurable range or scaling range of the range type*

* Within the scaling range: –5% to 105% of the scale value or –20000 to 30000 (the decimal

place is the same as the setting for the scale value)

Note

• Make sure that the leftmost value is less than the rightmost value for span and scale.
• Calibration correction cannot be set on ON/OFF input (DI), delta computation, and square

root computation channels. In addition, calibration correction cannot be used on all channels
if the bias function is enabled.

<Related Topcs> Enabling the calibration correction function: Section 7.15

Setting the calibration correction function: Section 7.23