7-42

7.25 Changng the Temperature Unt

This section explains the details of changing the temperature unit on the temperature
measurement channels.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).

Procedure

Changng the Settngs

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

2.

Hold down the

and

keys simultaneously for 3 seconds to enter Basic

Setting mode.

3.

Press the

key or

SHIFT +

key to select

Temperature and then press the

key.

4.

Set each item and press the key.

Use the

key or

SHIFT +

key to select values.

For the procedure on how to enter values or characters, see section 4.2.
If you press the

ESC key in the middle of the operation, the settings entered up to

then are cancelled, and the display returns to a higher level menu.

5.

When the

Settng complete screen appears, do either of the following:

To correct the setting, press the key.
If you are done, press the

ESC key.

Select the temperature unit.

The new setting takes effect.

ESC/?

Temperature

Setting complete

Basic=Temperature

Temp=°C

Applyng the Changes and Returnng to Operaton Mode

Press the

key or

SHIFT +

key to select

End and then press the key. Press

the

key to select

Store and then press the key. The changes are applied, and

the screen returns to Operation mode. If you select

Abort and press the key, the

changes are discarded, and the screen returns to Operation mode. Press the

ESC key to

return to the

Basc= screen.

Explanaton

Temp (Temperature)

The setting is applied to all temperature measurement channels.
°C: Use Celsius.
°F: Use Fahrenheit.