1.7 FAIL/Chart End Detecton and Output Functon

(/F1 Opton)

FAIL Output

When a failure occurs in the CPU of the recorder, a relay contact signal (1 signal) is
output. The relay is energized when the CPU is normal and de-energizes when a CPU
failure occurs. Therefore, relay output is carried out also when the power is turned OFF
(including a power failure). This relay behavior cannot be changed.

• FAIL Relay Output

NO

C

NC

Normal

NO

C

NC

Failure

NO

C

NC

Power-OFF

NO: Normally Opened, C: Common, NC: Normally Closed

Chart End Output

When the remaining amount of chart paper falls to approximately 2 cm, recording is
stopped, and the relay contact signal (1 relay) is output. The relay is energized in this
case. This relay behavior cannot be changed.

NO

C

NC

Normal

NO

C

NC

Chart end

NO

C

NC

Power-OFF

NO: Normally Opened, C: Common, NC: Normally Closed