Appendix 2 periodic printout interval, Appendix 2, Appendx 2 perodc prntout interval – Yokogawa µR10000 User Manual

Page 258

Advertising
background image

IM 04P01B01-01E

App-6

Appendx 2 Perodc Prntout Interval

The periodic printout function prints measured values and other data numerically at the
left side of the chart paper while performing trend recording.
The periodic printout interval varies depending on the chart speed and setup conditions.

Perodc Prntout Example of the Dot Model

Scale

Chart speed printout

Measured values/alarm status

Trend recording

Date/Time

Interval

Interval

Periodic printout

Periodic printout

Mar.31.04

08:30

1 0.001V

2 0.000V

3 0.000V

4 0.000V

5 0.000V

6 0.000V

-2.000

2.000

1CH

V

20mm/h_

Mar.31.04

10:00

1 0.001V

2 0.000V

3 0.000V

4 0.000V

5 0.000V

6 0.000V

-2.000

2.000

1CH

V

20mm/h_

Mar.31.04

11:30

1 0.001V

2 0.000V

3 0.000V

4 0.000V

5 0.000V

6 0.000V

-2.000

2.000

1CH

V

20mm/h_

The figure above is used to explain the printout contents. The actual printout and font are
different from those illustrated in the figure. The printout positions are also slightly different.

When the Interval Is Set to Auto

Printout is performed at the interval corresponding to the chart speed.

Pen Model

Dot Model

Chart Speed

Perodc Prntout

Interval

Chart Speed

Perodc Prntout

Interval

Up to 9 mm/h

Not printed

Up to 9 mm/h

Not printed

10 to 18 mm/h

8 hours

10 to 19 mm/h

8 hours

20 to 36 mm/h

4 hours

20 to 39 mm/h

4 hours

40 to 72 mm/h

2 hours

40 to 79 mm/h

2 hours

75 to 135 mm/h

1 hour

80 to 100 mm/h

1 hour

150 to 180 mm/h

30 minutes

101 mm/h or higher

Not printed

200 to 320 mm/h

20 minutes

360 to 1500 mm/h

10 minutes

1600 mm/h or higher

Not printed