6.13 Performng Calbraton Correcton (/CC1

Opton)

The values of the measuring input signal are corrected using specified segments, and
the results are used as measured values.
If you change the input range, set the correction point (measured value) and correction
values again.

Procedure

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

2.

Press the

key or

SHIFT +

key to select

Calbraton and then press the

key.

3.

Set each item and press the key.

Use the

key or

SHIFT +

key to select values.

For the procedure on how to enter values or characters, see section 4.2.
If you press the

ESC key in the middle of the operation, the settings entered up to

then are cancelled, and the display returns to a higher level menu.

4.

When the

Settng complete screen appears, do either of the following:

Press the key to set other channels.
If you are done, press the

ESC key.

5.

Hold down the

MENU

key for 3 seconds to return to Operation mode.

Set=Calibration

CH=01

01CH Calibration

Setting complete

Turn On or Off the calibration correction.

Set the first correction point

(input value).

Select the channel.

(Displays the following when

Calibration is On)

The new setting takes effect.

Yes: The setting is fixed.

No: The setting becomes invalid.

ESC/?

Calibration=On

1 datums= 0.000

Decision=Yes

Set the first correction value.
Likewise, set the second correction

point and correction value.

..

.

1 revise= 0.000