4.1 Run Modes

The recorder has three run modes.

Operation Mode

This mode is used for normal recording operation. The recorder enters this mode
when the power is turned ON. The operations that can be carried out in this mode are
described in chapter 3. The details on the data display setup screen are explained in
chapter 8.

Setting Mode

This mode is used to set the input range, alarms, chart speed, and other parameters.
These settings can be changed while recording is in progress. However, the input
range of measurement channels and the computing equation, unit, constant, and TLOG
setting of computation channels cannot be changed while computation (/M1 option) is in
progress. The operations that can be carried out in this mode are described in chapter 5
and 6.

Basic Setting Mode

This mode is used to set the basic specifications of the recorder such as the
thermocouple burnout detection function and the alarm output relay operation. This mode
cannot be entered while the recorder is recording or while computation is in progress
on the computation function (/M1 option). Measurement, recording, and alarm detection
cannot be carried out in this mode. The operations that can be carried out in this mode
are described in chapter 7.