background image

IM 04P01B01-01E

2-12

Wrng Procedure

Panel Mount Type

The power supply terminals and protective ground terminals are located on the rear
panel.

1.

Turn OFF the power switch on the recorder and open the power terminal cover.

Screw for fixing

the power terminal

cover in place

Power

terminal block

Power

terminal cover

Open

2.

Wire the power cord and the protective ground cord to the power supply terminals.

Use ring-tongue crimp-on lugs (designed for 4 mm screws).

Philips

screwdriver

Crimp-on lug

with insulation

sleeves

Power cord

Protective ground

cord

• Other than /P1 • /P1

(24 VDC/AC power supply)

L

N

L

N

DC/AC24V

+

-

3.

Close the power supply terminal cover and secure it with the screw. The proper

torque for tightening the screws is 0.6 N•m.

Portable Type (/H5x Opton)

1.

Check that the power switch to the recorder is turned OFF.

2.

Connect the plug on the accessory power cord to the power supply connector on

the rear panel.

100-240V~

50/60Hz

40V

A MAX

3.

Ensure that the power outlet to be used meets the conditions on the previous

page and that the voltage of the power supply is within the maximum voltage
rating of the power cord, then connect the other end of the power cord to the
power supply outlet.
The AC outlet must be of a three-prong type with a protective earth ground termi-
nal.

2.5 Power Supply Wrng