background image

2-7

IM 04P01B01-01E

Before

Us

ng

the
Recorder

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

Measurng Input Wrng

DC voltage

1-5V

ON/OFF

Extension leadwire

DC current

Shunt resistor

–/B

b

+/A

–/B

b

+/A

–/B

b

+/A

–/B

b

+/A

+

+

Thermocouple input

Resistance temperature detector input

DC voltage input, 1-5V input,

and ON/OFF input

DC current input

Leadwire resistance:

10 Ω max./wire. The resistance

of the three wires should be

equal.

Example: For a 4 to 20 mA

input, a shunt resistor of

250 Ω ± 0.1% can be used to

convert to 1-5V input.

A

B

b

Note

RTD input terminals A and B on the dot model are isolated on each channel. Terminal b is
shorted internally across all channels. However, for 3 legs isolated RTDs (/N2 option), input b is
also isolated for each channel.