App index 24 vdc/ac power supply operaton (/p1), Calbraton correcton (/cc1), Header prntout (/bt1) – Yokogawa µR10000 User Manual

Page 245

Advertising
background image

12-17

IM 04P01B01-01E

Spec

fcat
ons

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

24 VDC/AC Power Supply Operaton (/P1)

Item

Specfcatons

Rated supply voltage

24 VDC/AC

Allowable power supply voltage range

21.6 to 26.4 VDC/AC

Withstand voltage

1000 VAC at 50/60 Hz for one minute (between the power terminal and the ground terminal)

Rated power supply frequency

50/60 Hz (for AC operation)

Allowable power supply frequency range

50 Hz ± 2% or 60 Hz ± 2% (for AC operation)

Effects of power supply voltage fluctuation

The fluctuation in the measured values is within ±1 digit and that in the recording is within ±0.1% in

the recording span for the 21.6 to 26.4 VAC/DC range.

Effects of power supply frequency fluctuation

The fluctuation in the measured values and recording is ±(0.1% of rdg + 1 digit) at the rated

frequency ±2 Hz.

Rated power

25 VA (for DC) or 35 VA (for AC)

Power consumption

Model

24 VDC

24 VAC

Maxmum

1 to 4 pen models

Approx. 7 VA*

Approx. 13 VA*

Approx. 35 VA

6 dot model

Approx. 8 VA*

Approx. 13 VA*

Approx. 35 VA

* When balanced

Calbraton Correcton (/CC1)

Item

Specfcatons

Number of calibration points 2 to 16 (can be specified for each channel)
Correction mode

Revise or absolute value (can be specified for each channel)

Correctable range

DC input (DC voltage, TC, and RTD) and linear scaling (DC voltage, TC, RTD, and 1-5V)
(Not allowed on ON/OFF input (including linear scaling), delta computation, square root

computation, and computation channels)

Correction range

• (1

st

correction point + 1

st

correction value) < (2

nd

correction point + 2

nd

correction value) < ...

• Corrected value (revise value: correction point + correction value, absolute value: correction

value) within the measurable range and scaling range (–5% to 105% or –20000 to 30000) of each

input range

• Setting limits
Left span (0%) < Right span (100%)
Left scale (0%) < Right scale (100%)

Others

Calibration correction and bias functions cannot be used simultaneously (common to all channels)

Header Prntout (/BT1)

Item

Specfcatons

Printout function

When recording starts or stops, prints out the batch name, comment, date/time, and chart speed.

Messages can also be printed out including measured values*.

Printed items

Recording start/recording end printouts

Batch name: Batch number (up to 26 characters) and lot number (4 or 6 digit number, or Off)
Comment (up to 32 characters x 5 lines)
Chart speed
Date/time: Year, month, date, time, minute, second (date format depends on the date printout/

display format)
Printout On/Off selection: Batch name, chart speed, and date/time

Message printouts

In the 5 message formats, measured values*, strings (up to 16 characters), and date/time can be

arbitrarily combined and printed out (up to 35 characters total).

* Computed values can also be included on models with the /M1 option.

12.5 Specfcatons of Optonal Functons

Advertising