Effects of operatng condtons, Other specfcatons – Yokogawa µR10000 User Manual

Page 250

Advertising
background image

IM 04P01B01-01E

12-22

Effects of Operatng Condtons

Item

Specfcatons

Ambient temperature

With temperature variation of 10°C:

Measurement: ±(0.1% of rdg + 1 digit) or less

Recording:

Measurement fluctuation + ±0.2% of the recording span or less

* Excluding the error of reference junction compensation

Power supply fluctuation

With variation within 90 to 132 V and 180 to 264 VAC (50/60 Hz):

Measurement: ±1 digit or less

Recording:

±0.1% of the recording span or less

With variation of ±2 Hz from rated power frequency (power supply voltage 90 to 132, 180 to

264 VAC):

Measurement: ±(0.1% of rdg + 1 digit) or less

Recording:

Same as the measurement fluctuation

Magnetic field

AC (50/60 Hz) and DC 400 A/m fields:

Measurement: ±(0.1% of rdg 10 digit) or less

Recording:

±0.5% of the recording span or less

Input source resistance

DC voltage range

With variation of +1 kΩ:

Ranges of 200 mV or less:

Within ±10 µV

Ranges of 2V or more:

±0.1% of rdg

TC range

With variation of +1 kΩ: ±10 µV or less

RTD range

With variation of 10 Ω per wire (resistance of all three wires must be equal):

Measurement: ±(0.1% of rdg + 1 digit) or less

Recording:

Measurement fluctuation + ±0.1% of the recording span or less

With maximum difference of 40 mΩ between wires: Approx. 0.1°C (for Pt100)

Mounting position

With a backward tilt of 30° or less

Measurement: ±(0.1% of rdg + 1 digit) or less

Recording:

Indication fluctuation + ±0.1% of the recording span or less

Vibration

When a sinusoidal vibration along all three axis at a frequency between 10 to 60 Hz and an

acceleration of 0.2 m/s

2

is applied for 2 hours

Measurement: ±(0.1% of rdg + 1 digit) or less

Recording:

Indication fluctuation + ±0.1% of the recording span or less

Other Specfcatons

Item

Specfcatons

Clock

With a calendar function

Accuracy of clock

±100 ppm, excluding a delay (of 1 second, maximum) caused each time the power is turned on.

DST

The date/time for switching between standard time and DST can be specified.

Memory backup

A built-in lithium battery backs up the setup parameters and runs the clock

(battery life: approximately ten years at room temperature).

Key lock function

Password for releasing the key lock can be set.

Key lock items

RCD key, MENU key, DISP MENU key, FEED key, and function under the FUNC key (Alarm ACK,

computation start/stop, print, message, printout buffer clear, periodic printout reset, and pen or

ribbon cassette exchange)

Internal light

Lights up the recording area of the chart paper. Uses a dedicated LED.

Noise

Machine Noise Information Ordinance 3.GSGV, Jan 18, 1991

Maximum noise level: 60 dB (A) or less (complies with ISO7779)

12.6 General Specfcatons

Advertising