App index, Nose elmnaton from input sgnals – Yokogawa µR10000 User Manual

Page 21

Advertising
background image

1-5

IM 04P01B01-01E

Funct

onal
Explanat

on

and
Setup

Gu
de

For the procedure to set the functions, see section 1.10, “Function Setup Guide.”

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

Note

When using external reference junction compensation, set an appropriate reference junction
compensation voltage. For example, if the reference junction temperature of the external
reference compensation is T

0

°C, set the reference compensation junction voltage to the

thermoelectromotive force of the 0°C reference of T

0

°C.

Recorder

Example when using external reference junction compensation

External reference junction compensation

(Hold the contact point of the thermocouple

and copper wire at T

0

°

C)

Thermocouple

Copper wire

<Related Topcs> Setting the reference junction compensation function: Section 7.4

Nose Elmnaton from Input Sgnals

Flter and Movng Average

This function used to suppress the effects of noise that is riding on the signal. The pen
model and dot model are equipped with a filter function and a moving average function,
respectively. The function can be set for each measurement channel. However, it does
not operate on channels set to ON/OFF input.

• Flter (Pen Model)

The filter is a low-pass filter. The time constant can be set to 2 s, 5 s, or 10 s.

63.2% of the output value

2, 5, 10 s (time constant, the time it takes

to reach 63.2% of the output value)

Output response curve

(when using the filter)

Input

Filter result (output for a step input)

• Movng Average (Dot Model)

The average value of the m most recent values acquired at the scan interval is used

as the measured value of the channel. The number of moving-averaged data points
(m) can be set in the range 2 to 16. The figure below shows an example indicating
the operation of the buffer for the moving average computation when the number of
moving averaged data points is set to 5.

1
2
3
4
5

10.0 mV

5.0 mV
0.0 mV

–5.0 mV

–10.0 mV

0.0 mV

15.0 mV
10.0 mV

5.0 mV
0.0 mV

–5.0 mV

5.0 mV

10.0 mV
15.0 mV
10.0 mV

5.0 mV
0.0 mV

8.0 mV

Deleted

Deleted

Moving

average

Most recent data

Most recent data

Buffer data for the

nth sampling time

Buffer data for the

n+1th sampling time

Buffer data for the

n+2th sampling time

<Related Topcs> Setting the filter: Sections 7.12 and 6.2

Setting the moving average: Sections 7.11 and 6.3

1.2 Measurng Input Secton

Advertising