Recommended peripheral devices – IDEX Health & Science VS-606V7 Series User Manual

Page 153

Advertising
background image

153

Recommended Peripheral Devices

It is recommended that the following peripheral devices should be mounted between
the AC main circuit power supply and VS-606V7 input terminals R/L1, S/L2, and T/
L3.

MCCB (Molded-case circuit breaker)/fuse:
Be sure to connect it for wiring protection.

Magnetic contactor:
Mount a surge suppressor on the coil (refer to the table shown below).
When using a magnetic contactor to start and stop the inverter, do not exceed one
start per hour.

Recommended MCCB and magnetic contactor, and fuse

200V 3-phase

200V single-phase

400V 3-phase

VS-606V7 model

V7

∗ ∗

20P1

V7

∗ ∗

20P2

V7

∗ ∗

20P4

V7

∗ ∗

20P7

V7

∗ ∗

21P5

V7

∗ ∗

22P2

V7

∗ ∗

23P7

V7

∗ ∗

25P5

V7

∗ ∗

27P5

Capacity (kVA)

0.3

0.6

1.1

1.9

3.0

4.2

6.7

9.5

13.0

Rated Output Current (A)

0.8

1.6

3

5

8

11

17.5

25.0

33.0

MCCB type NF30
(MITSUBISHI)

5A

5A

5A

10A

20A

20A

30A

50A

60A

Magnetic contactor type HI
(YASKAWA CONTROL)

HI-7E

HI-7E

HI-7E

HI-7E HI-10-2E HI-10-2E

HI-20E

HI-30E HI-50E

Fuse (UL Class RK5)

5A

5A

5A

10A

20A

20A

30A

50A

60A

VS-606V7 model

V7

∗ ∗

B0P1

V7

∗ ∗

B0P2

V7

∗ ∗

B0P4

V7

∗ ∗

B0P7

V7

∗ ∗

B1P5

V7

∗ ∗

B2P2

V7

∗ ∗

B3P7

Capacity (kVA)

0.3

0.6

1.1

1.9

3.0

4.2

6.7

Rated Output Current (A)

0.8

1.6

3

5

8

11

17.5

MCCB type NF30, NF50
(MITSUBISHI)

5A

5A

10A

20A

20A

40A

50A

Magnetic contactor type HI
(YASKAWA CONTROL)

HI-7E

HI-7E

HI-7E HI-10-2E

HI-15E

HI-20E

HI-30E

Fuse (UL Class RK5)

5A

5A

10A

20A

20A

40A

50A

VS-606V7 model

V7

∗ ∗

40P2

V7

∗ ∗

40P4

V7

∗ ∗

40P7

V7

∗ ∗

41P5

V7

∗ ∗

42P2

V7

∗ ∗

43P0

V7

∗ ∗

44P0

V7

∗ ∗

45P5

V7

∗ ∗

47P5

Capacity (kVA)

0.9

1.4

2.6

3.7

4.2

5.5

7.0

11.0

14.0

Rated Output Current (A)

1.2

1.8

3.4

4.8

5.5

7.2

9.2

14.8

18.0

MCCB type NF30, NF50
(MITSUBISHI)

5A

5A

5A

10A

10A

20A

20A

30A

30A

Magnetic contactor type HI
(YASKAWA CONTROL)

HI-7E HI-7E HI-7E HI-10-2E HI-10-2E HI-10-2E HI-10-2E HI-20E HI-20E

Fuse (UL Class RK5)

5A

5A

5A

10A

10A

20A

20A

30A

30A

Advertising