Hand Probe P50

19

ninb63e1-f (1410)

4

Hand Probe P50

4.1

General Information

4.2

Description

Hand Probe P50 (Rigid Neck)

The Hand Probe P50 is a non-sniffing probe. Gas analysis takes place in the
replaceable sensor located in the tip of the probe.

For more information about the Hand Probe P50, refer to the Operating Manual and
Technical Reference Manual for Sensistor ISH2000.

4.3

Calibration

The probe needs to be calibrated for optimal accuracy. Make sure it is calibrated when
it is used for the first time. For more information about calibration, see on page 58.

Caution

Do not expose the probe to a hydrogen concentration greater than 0.1% when the
instrument is not operating, as this could damage or destroy the probe sensor.

Caution

Connection and disconnection of the sensor cable must be done with power OFF.
Sensor can be damaged if power is on.

1

Hydrogen Sensor

2

Indicator LED lamps

3

Function Button

4

Lamps

3

4

2

1

Notice

The probe can be ordered with a flexible neck. For more information, see
on page 99.