5 blockage test – INFICON Sensistor ILS500 Leak Detection System User Manual

Page 51

Advertising
background image

Recipes

51

ninb63e1-f (1410)

9.3.5

Blockage Test

Fill Setpoint

Desired tracer gas fill pressure.

Fill Timeout

Object will be rejected if Pressure Setpoint has not
been attained within this time.

Cancels the fill if the test object has a major leak,
opens, or if there are loose connections.

External Fill Regulation

If selected, this is the setpoint of fill pressure alarm.
Internal pressure regulation is disengaged and
pressure will be that of the gas supply line. ILS500
checks that fill pressure is above Pressure Setpoint
before proceeding to gas test step.

Pressure Unit

Select desired unit.

Notice

This test can only be performed if both test ports are used and connected
on either side of the possible blockage.

Blockage Test

Blockage Test Pressure

Blockage Test Time

0.50

2.0

bar

s

Blockage Test Pressure

Minimum pressure to be attained at Test Port 2
during Blockage Test time.

Blockage Test Time

Time within which Blockage Test Pressure must be
attained at Test Port 2.