INFICON Sensistor ILS500 Leak Detection System User Manual
Page 95

Technical Data
95
ninb63e1-f (1410)
For further information about APC system refer to the Technical Reference Manual for
Sensistor ISH2000.
14.4.8
Power Input
14.4.9
Safety Interface
18
DET_WAIT
OUT
0.5 A
Output of ISH2000 APC system.
19
OUT_0
OUT
0.5 A
Output of ISH2000 APC system.
20
OUT_1
OUT
0.5 A
Output of ISH2000 APC system.
21
OUT_2
OUT
0.5 A
Output of ISH2000 APC system.
22
OUT_3
OUT
0.5 A
Output of ISH2000 APC system.
23
OUT_4
OUT
0.5 A
Output of ISH2000 APC system.
24
OUT_5
OUT
0.5 A
Output of ISH2000 APC system.
25
24 VDC OUT
OUT
0.5 A
Supply for switches etc. connecting to
inputs. Do not feed 24 V here.
Pin
Signal
Type
Load
Comment
Specification
AC mains voltage
110-240 V 50/60Hz.
AC mains current
Typically 1 A (2 A pulse at power on).
Caution
Risk assessment is the sole duty of the user of the ILS500
Connector:
6 pin male Weidmüller, Omnimate BL3.5. Mating screw terminal
included.
Purpose:
Emergency stop interface.