background image

50

Recipes

ninb63e1-f (1410)

Pressure Decay Test

9.3.4

Tracer Gas Filling

Pressure Stabilisation Time

Pressure Decay Test Time

Pressure Decay Limit

Gas Locate if gross leak

5.0

5.0

0.10

s

s

bar

Gas Pressure Decay Test

Pressure Stabilisation Time

Delay time before Pressure Decay test begins.

Pressure Decay Test Time

Time during which pressure drop is recorded.

Pressure Decay Limit

Allowed pressure drop during test time.

Gas Locate if gross leak

If there is a need to locate a leak with a Hand
Probe even if pre evacuation fails. ILS500 will fill to
chosen pressure in Locating Pressure under
Tracer Gas Test.

Warning

The ILS500 must never be introduced to pressures higher than that approved for the
object to be tested and never beyond the ILS500 specification.

Notice

Ensure that the test object has time to become filled before Fill Time Out
expires. In particular long narrow objects, as pipes, may need long filling
time.

Tracer Gas Filling

Fill Setpoint

Fill Timeout

External Fill Regulation

Pressure Unit

5.00

10.0

bar

bar

s