background image

Troubleshooting

65

ninb63e1-f (1410)

Venturi Pump

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

-0.88

0.00

bar

bar

Venturi Pump

Max Vacuum OK

Checking max vacuum of Venturi pump.

Possible results:

Max Vacuum OK

Poor Max Vacuum

Poor max vacuum can result in:

Failed pre-evacuation

Slower evacuation

Manifold Tightness (gross)

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

-0.82 bar

bar

Manifold Tightness

No Leakage from Outside

0.00

The overall tightness of the manifold is tested using the vacuum raise method.

Possible results:

No Leakage from Outside

Leakage from Outside

Leaks in the manifold can result in:

False vacuum decay rejects

Increased gas consumption

Minor external leakage will be found later during the

gas test step.