Calibration

ninb63e1-f (1410)

10

Calibration

10.1

About Calibration

It is important to have a correct calibration when measuring the size of a leak in
Measure Mode and Combined Mode.

There are two ways to calibrate the probe:

Calibration gas (recommended)
Has a known concentration of hydrogen (10 ppm recommended).

Calibration leak
Has a fixed leak rate (flow or g/y).

Both methods take less than 2 minutes to perform.

Calibration can be performed automatically for the Active Probes. For Hand Probes the
ILS500 will display a reminder when the system is due for calibration.

10.1.1

Required Equipment

Calibration Gas (recommended) or Calibration Leak

Relevant Certificate

For more information, see on page 99.

10.2

Calibration Overview

Click Settings >> Calibration to get to the three Calibration Setup menus.

Notice

If calibration fails you can still use the instrument. Last valid calibration
parameters will be used. However, you should check that the instrument
reacts to the gas.

Calibration Settings

Calibration Value

Calibration Unit

Leak Gas

cc/s

Air

2.2E-05