4 replacing the venturi pump – INFICON Sensistor ILS500 Leak Detection System User Manual

Page 79

Advertising
background image

Maintenance Instructions

79

ninb63e1-f (1410)

12.3.4

Replacing the Venturi Pump

1

Remove the exhaust hose from the barbed hose fitting

2

Unscrew and remove the barbed hose fitting and the plastic washer.

3

Use a 4 mm Allen key to remove the four screws holding the Venturi pump.

4

Remove the o-ring under the Venturi. Remove the hose from Venturi inlet.

Push hose into connector and press orange ring down to release hose, then pull
hose out.

5

Remove the hose fitting from the Venturi.

Advertising