INFICON Sensistor ILS500 Leak Detection System User Manual

Page 59

Advertising
background image

Calibration

59

ninb63e1-f (1410)

Calibration Settings

Calibrate:

At Startup

After Recipe change

Every

test

Calibration Repeat Pause

50

30

s

Calibration Settings

Cal. Leak in test cycle

Prevent Start

Automatic (Active Probe only)

Max. Attempts

5

Cal. Leak Pressure

Set Cal. Leak Pressure

1.00 bar

Calibration Value

Set to the same as stated on the calibration
certificate issued for the leak or gas.

Calibration Unit

Set to the same as stated on the calibration
certificate issued for the leak or gas.

Leak Gas

Set the gas flow for which the Calibration Leak is
defined.

Notice

For more information, refer to the manuals for Sensistor ISH2000.

Calibrate:

At Startup

Calibration is performed or requested whenever
the power is switched on.

After Recipe Change

Calibration is requested every time another recipe
is selected.

Every “XX” Test

Set number of tests between calibration requests.

Calibration Repeat Pause

Set minimum time between calibrations (recovery
time for sensor).
The unit will signal the interval through a pulsating
sound indication together with the text "Calibrate!".

Notice

The best way to find optimal calibration for an application, is to test
different intervals and learn at what interval calibration is optimal for the
specific case.