11 troubleshooting, 1 fault symptoms, 2 perform hardware test – INFICON Sensistor ILS500 Leak Detection System User Manual

Page 62: Troubleshooting, Fault symptoms, Perform hardware test, 1 fault symptoms 11.2 perform hardware test

Advertising
background image

62

Troubleshooting

ninb63e1-f (1410)

11

Troubleshooting

11.1

Fault Symptoms

11.2

Perform Hardware Test

Fault Symptom

Fault

Measures

Evacuation Failed

Failed to reach vacuum
within the specified time.

Large leak on Test Object
or connections.

Check the compressed air
supply.

Gas Fill Failed

Failed to fill to the right
pressure within the
specified time.

Large leak on Test Object
or connections.

Check the incoming gas
pressure.

Gas Refill Failed

Failed to refill the object.

Large leak on Test Object
or connections.

Check the incoming gas
pressure.

Gas Evac Failed

Failed to reach vacuum
within the specified time.

Detector Signal

Detector is busy signaling,
wait until finished.

Check the fresh air supply.

ISH2000 APC Driver Error

ISH2000 failed to start
measurement during test.

Check that ISH2000 APC is
correctly configured.

Detector Not Configured!

Hardware settings are not
correctly set.

Go to Hardware Setup and
set the correct hardware.

HW Error During Test

Serious error has occurred
during test.

Check external equipment,
e.g Active Probe.

Test Timeout

Maximum time for the test
was exceeded.

Check that time is correctly
set.

Notice

Before performing the hardware test, carefully check that your tracer gas
and compressed air feed pressures are correct. Wrongly set pressure can
cause erroneous test results.