3 gross leak tests – INFICON Sensistor ILS500 Leak Detection System User Manual

Page 49

Advertising
background image

Recipes

49

ninb63e1-f (1410)

9.3.3

Gross Leak Tests

Vacuum Decay Test

Gross Leak Test

Evacuation Timeout

Vacuum Decay Test
- before gas test

Pressure Decay Test
- during gas test

10.0 s

Evacuation Timeout

Object will be rejected if Pre Evacuation Setpoint is
not attained within time set.

Vacuum Decay Test

If to be included in the test sequence, select the
box and click the blue button to enter Pressure
Decay Test setup menu (see below).

Pressure Decay Test

If to be included in the test sequence, select the
box and click the blue button to enter Pressure
Decay Test setup menu (see below).

Vacuum Stabilisation Time

Vacuum Decay Test Time

Vacuum Decay Limit

Gas Locate if gross leak

5.0

5.0

0.10

s

s

bar

Vacuum Decay Test

Vacuum Stabilisation Time

Delay time before Vacuum Decay test begins.

Vacuum Decay Test Time

Time during which pressure rise is recorded.

Vacuum Decay Limit

Allowed pressure rise during test time.

Gas Locate if gross leak

If there is a need to locate a leak with a Hand
Probe even if pre evacuation fails. ILS500 will fill to
chosen pressure in Locating Pressure under
Tracer Gas Test.