INFICON Sensistor ILS500 Leak Detection System User Manual
Page 96
Advertising
96
Technical Data
ninb63e1-f (1410)
14.4.10 Connection Port
14.4.11 Leak Detector
Pin
Signal
Type
Load
Comment
1
+24 VDC
SUPPLY
2.5 A
2
AUX1
-
+/-1-5 A*
Terminal 1 of safe relay contacts for
auxiliary external use.
3
AUX2
-
+/-1-5 A*
Terminal 2of safe relay contacts for
auxiliary external use.
4
Not used
-
-
-
5
SAFESPLY**
SUPPLY
-2.5 A
24 VDC supply from EXTERNAL
emergency stop circuitry.
6
COM
GND
1.0 A
Common GND.
*
250 VAC 5 A cosj =1
30 VDC 5 A L/R = 0 ms
240 VAC 2A cosj = 0.3
24 VDC 1A L/R = 48 ms
** SAFESPLY feeds risk associated loads inside the ILS500. These include all gas
and tooling valves.
Connector:
Probe connector.
Purpose:
For connection of probe.
Cable:
C21 Probe Cable.
Advertising