INFICON Sensistor ILS500 Leak Detection System User Manual

Page 91

Advertising
background image

Technical Data

91

ninb63e1-f (1410)

14.4.2

Input 1 (Optional)

14.4.3

Input 2 (Optional)

Connector:

5 pin male Weidmüller, Omnimate BL3.5. Mating screw terminal
included.

Purpose:

Options port 1. Optional analogue or digital input (not supported by
std software).

Pin

Signal

Type

Load

Comment

1

+24 VDC

SUPPLY

250 mA

Option supply.

2

VIN1

IN

-60 mA

Voltage input:Digital 24 VDC or
analogue 0-10 VDC.

3

IIN1

IN

+/-30 mA

Current input: 0-20 mA.

4

COM1

IN

-250 mA

Signal common (GND).

5

COM/SHLD

GND

+/-30 mA

Shield/screed connection.

Connector:

5 pin male Weidmüller, Omnimate BL3.5. Mating screw terminal
included.

Purpose:

Options port 2. Used for “Active Holder for Hand Probe” (90630).

Pin

Signal

Type

Load

Comment

1

+24 VDC

SUPPLY

250 mA

Option supply.

2

VIN2

IN

-60 mA

Voltage input:Digital 24 VDC or
analogue 0-10 VDC.

3

IIN2

IN

+/-30 mA

Current input: 0-20 mA.

4

COM2

IN

-250 mA

Signal common (GND).

Advertising