INFICON Sensistor ILS500 Leak Detection System User Manual

Page 66

Advertising
background image

66

Troubleshooting

ninb63e1-f (1410)

Vacuum Sensor Valve

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

0.00 bar

bar

Vacuum Sensor Valve

Valve works!

0.00

This checks that the valve shuts to protect vacuum sensor before filling.

Possible results:

Valve works

Faulty!

Malfunction can result in:

Damage to vacuum sensor

Failed pre-evacuation

Tracer Gas Fill Valve

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

0.00 bar

bar

Tracer Gas Fill Valve

No Internal Leakage

0.00

The step tests the gas fill valve for internal leakage by registering pressure rise behind
the valve.

Possible results:

No Internal Leakage

Internal Leakage

Internal leakage can result in:

Erroneous pressure decay results

False vacuum decay rejects increased gas consumption

Advertising