Troubleshooting

ninb63e1-f (1410)

External Gas Leaks

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

0.00 bar

bar

Check for Leaks with Hand Probe

0.50

The ILS500 is now prepared for a manual test for external leakage. Use the hand probe
to check for leakage

1

Start by checking all connections between the ILS500 and your test object. Follow
each test line carefully and check every joint.

2

Proceed to check around the gas valves and manifold inside the ILS500.

Manifold Tightness (gross)

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

0.00 bar

bar

Manifold Tightness

0.46

Internal Leakage

The overall tightness of the manifold is tested using the pressure decay method. This is
a complement to the gas test, revealing leakage out, through the exhaust etc.

Possible results:

No Internal Leakage

Internal Leakage

Internal leakage can result in:

False pressure and vacuum decay rejects

Increased tracer gas consumption