INFICON Sensistor ILS500 Leak Detection System User Manual
Page 68

68
Troubleshooting
ninb63e1-f (1410)
External Gas Leaks
Hardware Test
CONTINUE
STOP
Vacuum:
Pressure:
0.00 bar
bar
Check for Leaks with Hand Probe
0.50
The ILS500 is now prepared for a manual test for external leakage. Use the hand probe
to check for leakage
1
Start by checking all connections between the ILS500 and your test object. Follow
each test line carefully and check every joint.
2
Proceed to check around the gas valves and manifold inside the ILS500.
Manifold Tightness (gross)
Hardware Test
CONTINUE
STOP
Vacuum:
Pressure:
0.00 bar
bar
Manifold Tightness
0.46
Internal Leakage
The overall tightness of the manifold is tested using the pressure decay method. This is
a complement to the gas test, revealing leakage out, through the exhaust etc.
Possible results:
•
No Internal Leakage
•
Internal Leakage
Internal leakage can result in:
•
False pressure and vacuum decay rejects
•
Increased tracer gas consumption