INFICON Sensistor ILS500 Leak Detection System User Manual

Page 64

Advertising
background image

64

Troubleshooting

ninb63e1-f (1410)

Pressure and Vacuum Sensors

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

0.00

0.00

Pressure and Vacuum Sensors

Zero Points are OK

bar

bar

Zero points of pressure and vacuum sensors are tested.

Possible results:

Zero Points are OK

Vacuum Zero Point not OK

Offset zero point can result in:

Incorrect gas filling

Erroneous vacuum or pressure decay results

Evacuation Valve

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

-0.03

0.00

bar

bar

Evacuation Valve

No Internal Leakage

Evacuation valve is checked for internal leakage.

Possible results:

No Internal Leakage

Internal Leakage

Internal leakage can result in:

False vacuum decay rejects

Increased tracer gas consumption

Advertising