Troubleshooting

ninb63e1-f (1410)

11.2.1

Hardware Error Messages

Error Message

Reason for Error

Corrective Action*

Detector Power Off

No power to the Leak
Detector.

Check the power cable to
the detector (internal or
external).

Detector Error + Check
Probe and Cable**

Probe cable disconnected. Connect cable.

Detector Error + Check
Sensor, Voltage Error**

Gas sensor damaged.

Replace probe or sensor.

Detector Error + Error
[Driver name]**

Active Probe error.

Consult Active Probe
Manual.

Hardware Error

Vacuum Sensor Error

No power to vacuum
sensor.

Check cable to sensor.

Sensor not connected to
AD.

Check connection to AD.

Damaged vacuum sensor. Send in for repair.

Hardware Error

Pressure Sensor Error

No power to pressure
sensor.

Check cable to sensor.

Sensor not connected to
AD.

Check connection to AD.

Damaged pressure sensor. Replace sensor.

Analog Inputs Power Off

No power to AD module.

Check power cable on left
side of AD module.

Emergency Stop Activated

Emergency Stop not
Reset.

Pull out emergency Stop
Button to reset.

ISH2000 Comm. Error

ISH2000 printer mode
turned off manually.

Restart system.

ISH2000 serial cable
disconnected.

Connect cable

(internal or external).

*

Contact your supplier if the suggested action does not clear the error.