2 pneumatic specifications, Pneumatic specifications – INFICON Sensistor ILS500 Leak Detection System User Manual

Page 86

Advertising
background image

86

Technical Data

ninb63e1-f (1410)

14.2

Pneumatic Specifications

Data rate

1200-115200 baud

Data bits

8

Stop bits

1

Parity

None

Flow control

None

Communication Ports

Compressed Air Supply

Pressure

Std Model

0.35–0.7 MPa

(3.4–6.9 bar)

(50–100 psi)

Reduced vacuum capacity below:
0.5 MPa

(4.8 bar)

(70 psi)

HP model

0.5–0.7 MPa

(4.8–6.9 bar)

(70–100 psi)

Peak Consumption
at 6 bar (87 psi)

240 l/min (508 SCFH)

Quality

Oil free and filtered to 5

μ

m

Dew point

Max 10°C (50°F)

Tracer Gas Supply

Composition

95% N

2

and 5% H

2

Pressure

Std Model

0.005–1.0 MPa

(0.05–10.0 bar)

(0.72–145 psi)

HP model

0.02–4.5 MPa

(0.2–45.0 Bar)

(3–652 psi)

Quality

Industrial grade purity (>95% purity)

Exhaust

Capacity in Exhaust Duct

Min 30 m

3

/h (1000 SCFH)

Dimensions of Hose Leading to
Duct

ID 25 mm (1 in.)